Validation of heart rate monitor-based predictions of oxygen uptake and energy expenditure
Author(s)
Montgomery, Paul G
Green, Daniel J
Etxebarria, Naroa
Pyne, David B
Saunders, Philo U
Minahan, Clare L
Griffith University Author(s)
Year published
2009
Metadata
Show full item recordAbstract
Validation of heart rate monitor-based predictions of oxygen uptake and energy expenditure. J Strength Cond Res 23(5): 1489-1495, 2009-To validate o2 and energy expenditure predictions by the Suunto heart rate (HR) system against a first principle gas analysis system, well-trained male (n = 10, age 29.8 ± 4.3 years, o2 65.9 ± 9.7 ml·kg− 1·min− 1) and female (n = 7, 25.6 ± 3.6 years, 57.0 ± 4.2 ml·kg− 1·min− 1) runners completed a 2-stage incremental running test to establish submaximal and maximal oxygen uptake values. Metabolic cart values were used as the criterion measure of o2 and energy expenditure (kJ) and compared ...
View more >Validation of heart rate monitor-based predictions of oxygen uptake and energy expenditure. J Strength Cond Res 23(5): 1489-1495, 2009-To validate o2 and energy expenditure predictions by the Suunto heart rate (HR) system against a first principle gas analysis system, well-trained male (n = 10, age 29.8 ± 4.3 years, o2 65.9 ± 9.7 ml·kg− 1·min− 1) and female (n = 7, 25.6 ± 3.6 years, 57.0 ± 4.2 ml·kg− 1·min− 1) runners completed a 2-stage incremental running test to establish submaximal and maximal oxygen uptake values. Metabolic cart values were used as the criterion measure of o2 and energy expenditure (kJ) and compared with the predicted values from the Suunto software. The 3 levels of software analysis for the Suunto system were basic personal information (BI), BI + measured maximal HR (BIhr), and BIhr + measured o2 (BIhr + v). Comparisons were analyzed using linear regression to determine the standard error of the estimate (SEE). Eight subjects repeated the trial within 7 days to determine reliability (typical error [TE]). The SEEs for oxygen consumption via BI, BIhr, and BIhr + v were 2.6, 2.8, and 2.6 ml·kg− 1·min− 1, respectively, with corresponding percent coefficient of variation (%CV) of 6.0, 6.5, and 6.0. The bias compared with the criterion o2 decreased from −6.3 for BI, −2.5 for BIhr, to −0.9% for BIhr + v. The SEE of energy expenditure improved from BI (6.74 kJ) to BIhr (6.56) and BIhr + v (6.14) with corresponding %CV of 13.6, 12.2, and 12.7. The TE values for o2 were ∼0.60 ml·kg− 1·min− 1 and ∼2 kJ for energy expenditure. The %CV for o2 and energy expenditure was ∼1 to 4%. Although reliable, basic HR-based estimations of o2 and energy expenditure from the Suunto system underestimated o2 and energy expenditure by ∼6 and 13%, respectively. However, estimation can be improved when maximal HR and o2 values are added to the software analysis.
View less >
View more >Validation of heart rate monitor-based predictions of oxygen uptake and energy expenditure. J Strength Cond Res 23(5): 1489-1495, 2009-To validate o2 and energy expenditure predictions by the Suunto heart rate (HR) system against a first principle gas analysis system, well-trained male (n = 10, age 29.8 ± 4.3 years, o2 65.9 ± 9.7 ml·kg− 1·min− 1) and female (n = 7, 25.6 ± 3.6 years, 57.0 ± 4.2 ml·kg− 1·min− 1) runners completed a 2-stage incremental running test to establish submaximal and maximal oxygen uptake values. Metabolic cart values were used as the criterion measure of o2 and energy expenditure (kJ) and compared with the predicted values from the Suunto software. The 3 levels of software analysis for the Suunto system were basic personal information (BI), BI + measured maximal HR (BIhr), and BIhr + measured o2 (BIhr + v). Comparisons were analyzed using linear regression to determine the standard error of the estimate (SEE). Eight subjects repeated the trial within 7 days to determine reliability (typical error [TE]). The SEEs for oxygen consumption via BI, BIhr, and BIhr + v were 2.6, 2.8, and 2.6 ml·kg− 1·min− 1, respectively, with corresponding percent coefficient of variation (%CV) of 6.0, 6.5, and 6.0. The bias compared with the criterion o2 decreased from −6.3 for BI, −2.5 for BIhr, to −0.9% for BIhr + v. The SEE of energy expenditure improved from BI (6.74 kJ) to BIhr (6.56) and BIhr + v (6.14) with corresponding %CV of 13.6, 12.2, and 12.7. The TE values for o2 were ∼0.60 ml·kg− 1·min− 1 and ∼2 kJ for energy expenditure. The %CV for o2 and energy expenditure was ∼1 to 4%. Although reliable, basic HR-based estimations of o2 and energy expenditure from the Suunto system underestimated o2 and energy expenditure by ∼6 and 13%, respectively. However, estimation can be improved when maximal HR and o2 values are added to the software analysis.
View less >
Journal Title
Journal of Strength and Conditioning Research
Volume
23
Issue
5
Subject
Sports science and exercise
Exercise physiology
Sports science and exercise not elsewhere classified
Medical physiology